Description
Temperature Sensor |
|||||
Range |
-20° to 70°C(-4° to 158°F)
|
||||
Accuracy |
±0.21°C from 0° to 50°C, see manual for plot(±0.38°F from 32° to 122°F, see manual for plot)
|
||||
Resolution |
0.024°C at 25°C, see manual for plot(0.04°F at 77°F, see manual for plot)
|
||||
Drift |
<0.1°C per year(0.18°F per year)
|
||||
RH Sensor* |
|||||
Range |
1% to 90% (non-condensing)
|
||||
Accuracy |
±2% from 20% to 80% typical to a maximum of ±4.5% including hysteresis at 25°C (77°F); below 20% and above 80% ±6% typical
|
||||
Resolution |
0.01%
|
||||
Drift |
<1% per year typical
|
||||
Response Time |
|||||
Temperature
|
7:30 minutes in air moving 1 m/s(7:30 minutes in air moving 2.2 mph)
|
||||
PH |
20 seconds to 90% in airflow of 1 m/s(20 seconds to 90% in airflow of 2.2 mph)
|
||||
Logger |
|||||
Radio Power |
1 mW(0 dBm)
|
||||
Transmission Range |
Approximately 30.5 m line-of-sight(Approximately 100 ft line-of-sight)
|
||||
Wireless Data Standard |
Bluetooth Low Energy (Bluetooth Smart)
|
||||
Logger Operating Range |
-20° to 70°C; 0 to 95% RH (non-condensing)
(-4° to 158°F; 0 to 95% RH (non-condensing))
|
||||
Logging Rate |
1 second to 18 hours
|
||||
Logging Modes |
Fixed interval (normal, statistics) or burst
|
||||
Memory Modes |
Wrap when full or stop when full
|
||||
Start Modes |
Immediate, push button, date & time, or next interval
|
||||
Stop Modes |
When memory full, push button, date & time, or after a set logging period
|
||||
Restart Mode |
Push button
|
||||
Time Accuracy |
±1 minute per month at 25°C(±1 minute per month at 77°F)
|
||||
Battery Life |
1 year, typical with logging interval of 1 minute. Faster logging and/or statistics sampling intervals, entering burst logging mode, and remaining connected with the app will impact battery life. Excessive readouts, audible alarms, and paging all impact battery life. Visual alarms and other events can have a marginal impact on battery life.
|
||||
Battery Type |
Two AAA 1.5 V alkaline batteries, user replaceable
|
||||
Full Memory Download Time |
Approximately 60 seconds; may take longer the further the device is from the logger
|
||||
LCD |
LCD is visible from 0° to 50°C;(LCD is visible from 32° to 122°F;)
the LCD may react slowly or go blank in temperatures outside this range |
||||
Size |
3.66 x 8.48 x 2.29 cm(1.44 x 3.34 x 0.9 in.)
|
||||
Weight |
56 g(1.98 oz)
|
||||
Environmental Rating | IP50 |