Description
Pressure (Absolute) and Water Level Measurements MX2001-01-SS-S and MX2001-01-Ti-S |
|||||
Operating Range | 0 to 207 kPa; approximately 0 to 9 m of water depth at sea level, or 0 to 12 m of water at 3,000 m of altitude | (0 to 30 psia; approximately 0 to 30 ft of water depth at sea level, or 0 to 40 ft of water at 10,000 ft of altitude) | |||
Factory Calibrated Range |
69 to 207 kPa, 0° to 40°C
(10 to 30 psia, 32° to 104°F)
|
||||
Burst Pressure |
310 kPa or 18 m depth(45 psia or 60 ft depth)
|
||||
Water Level Accuracy* |
Typical error: ±0.05% FS, 0.5 cm water
(Typical error: ±0.05% FS, 0.015 ft water)
Maximum error: ±0.1% FS, 1.0 cm water
(Maximum error: ±0.1% FS, 0.03 ft water)
|
||||
Raw Pressure Accuracy** |
±0.3% FS, 0.62 kPa maximum error(±0.3% FS, 0.09 psi maximum error)
|
||||
Resolution |
<0.02 kPa, 0.21 cm water(<0.003 psi, 0.007 ft water)
|
||||
Pressure Response Time (90%)*** | <1 second at a stable temperature | ||||
Pressure (Absolute) and Water Level Measurements MX2001-02-SS-S |
|||||
Operating Range |
0 to 400 kPa;(0 to 58 psia);
approximately 0 to 30.6 m of water depth at sea level, or 0 to 33.6 m of water at 3,000 m of altitude
|
(approximately 0 to 100 ft of water depth at sea level, or 0 to 111 ft of water at 10,000 ft of altitude) | |||
Factory Calibrated Range |
69 to 400 kPa, 0° to 40°C(10 to 58 psia, 32° to 104°F)
|
||||
Burst Pressure |
500 kPa or 40.8 m depth(72.5 psia or 134 ft depth)
|
||||
Water Level Accuracy* |
Typical error: ±0.05% FS, 1.5 cm water(Typical error: ±0.05% FS, 0.05 ft water)
Maximum error: ±0.1% FS, 3.0 cm water(Maximum error: ±0.1% FS, 0.1 ft water)
|
||||
Raw Pressure Accuracy** |
±0.3% FS, 1.20 kPa maximum error(±0.3% FS, 0.17 psi maximum error)
|
||||
Resolution |
<0.04 kPa, 0.41 cm water(<0.006 psi, 0.013 ft water)
|
||||
Pressure Response Time (90%)*** | <1 second at a stable temperature | ||||
Pressure (Absolute) and Water Level Measurements MX2001-03-SS-S |
|||||
Operating Range |
0 to 850 kPa; approximately 0 to 76.5 m of water depth at sea level, or 0 to 79.5 m of water at 3,000 m of altitude
|
(0 to 123.3 psia; approximately 0 to 251 ft of water depth at sea level, or 0 to 262 ft of water at 10,000 ft of altitude) | |||
Factory Calibrated Range |
69 to 400 kPa, 0° to 40°C(10 to 58 psia, 32° to 104°F)
|
||||
Burst Pressure |
500 kPa or 40.8 m depth(72.5 psia or 134 ft depth)
|
||||
Water Level Accuracy* |
Typical error: ±0.05% FS, 1.5 cm water(Typical error: ±0.05% FS, 0.05 ft water)
Maximum error: ±0.1% FS, 3.0 cm water(Maximum error: ±0.1% FS, 0.1 ft water)
|
||||
Raw Pressure Accuracy** |
±0.3% FS, 1.20 kPa maximum error(±0.3% FS, 0.17 psi maximum error)
|
||||
Resolution |
<0.04 kPa, 0.41 cm water(<0.006 psi, 0.013 ft water)
|
||||
Pressure Response Time (90%)*** | <1 second at a stable temperature | ||||
Pressure (Absolute) and Water Level Measurements MX2001-04-SS-S and MX2001-04-Ti- |
|||||
Operating Range |
0 to 145 kPa; approximately 0 to 4 m of water depth at sea level, or 0 to 7 m of water at 3,000 m of altitude
|
(0 to 21 psia; approximately 0 to 13 ft of water depth at sea level, or 0 to 23 ft of water at 10,000 ft of altitude) | |||
Factory Calibrated Range |
69 to 145 kPa, 0° to 40°C(10 to 21 psia, 32° to 104°F)
|
||||
Burst Pressure |
310 kPa or 18 m depth(45 psia or 60 ft depth)
|
||||
Water Level Accuracy* |
Typical error: ±0.075% FS, 0.3 cm water(Typical error: ±0.075% FS, 0.01 ft water)
Maximum error: ±0.15% FS, 0.6 cm water(Maximum error: ±0.15% FS, 0.02 ft water)
|
||||
Raw Pressure Accuracy** |
±0.3% FS, 0.43 kPa maximum error(±0.3% FS, 0.063 psi maximum error)
|
||||
Resolution |
<0.014 kPa, 0.14 cm water(<0.002 psi, 0.005 ft water)
|
||||
Pressure Response Time (90%)*** | <1 second at a stable temperature | ||||
Barometric Pressure (MX2001-TOP) |
|||||
Operating Range |
0 to 400 kPa;(0 to 58 psia);
|
||||
Factory Calibrated Range |
-20 to 50°C(-4 to 122°F)
|
||||
Burst Pressure |
±0.2 kPa over full temperature range at fixed pressure; maximum error ±0.5% FS
(±0.029 psi over full temperature range at fixed pressure; maximum error ±0.5% FS)
|
||||
Water Level Accuracy* |
Typical error: ±0.075% FS, 0.3 cm water(Typical error: ±0.075% FS, 0.01 ft water)
Maximum error: ±0.15% FS, 0.6 cm water(Maximum error: ±0.15% FS, 0.02 ft water)
|
||||
Resolution |
<0.01 kPa(<0.0015 psi)
|
||||
Response Time |
<1 second at stable temperature
|
||||
Stability (Drift)* |
<0.01 kPa per year(<0.0015 psi per year)
|
||||
Temperature Measurements (All Sensor End Models MX2001-0x-SS-S and MX2001-0x-Ti-S) |
|||||
Operating Range |
-20° to 50°C(-4° to 122°F)
|
||||
Accuracy |
±0.44°C from 0° to 50°C, see manual for plot(±0.79°F from 32° to 122°F, see manual for plot)
|
||||
Resolution |
0.1°C at 25°C, see manual for plot(0.18°F at 77°F, see manual for plot)
|
||||
Response Time (90%) |
5 minutes in water (typical)
|
||||
Stability (Drift) |
0.1°C per year(0.18°F per year)
|
||||
Logger | |||||
Operating Range |
-20° to 50°C(-4° to 122°F)
|
||||
Radio Power |
1 mW(0 dBm)
|
||||
Transmission Range |
Approximately 30.5 m line-of-sight
(Approximately 100 ft line-of-sight)
|
||||
Wireless Data Standard |
Bluetooth Low Energy (Bluetooth Smart®)
|
||||
logging Rate |
1 second to 18 hours
|
||||
Logging Modes |
Fixed interval, multiple intervals with up to 8 user-defined logging intervals and durations, or burst
|
||||
Memory Modes |
Wrap when full or stop when full
|
||||
Start Modes |
Immediate, date & time, or next interval
|
||||
Stop Modes |
When memory full, date & time, or after a set logging period
|
||||
Time Accuracy |
±1 minute per month 0° to 50°C
(±1 minute per month 32° to 122°F)
|
||||
Battery |
Two AA, 1.5 V alkaline batteries, user replaceable
|
||||
Battery Life |
1 year, typical with logging interval of 1 minute. Faster logging and/or statistics sampling intervals, entering burst logging mode, excessive readouts, checking of Full Status Details, and remaining connected with the app will impact battery life.
|
||||
Memory |
256 KB memory (30,000 sets of measurements)
|
||||
Full Memory Download Time |
Approximately 2 minutes; may take longer the further the device is from the top end of the logger
|
||||
Dimensions |
Top end (MX2001-TOP): 2.54 cm (1.0 inches) diameter, 28.9 cm (11.4 inches) length; mounting hole 7.6 mm (0.3 inches) diameter
Sensor end (MX2001-0x-SS-S and MX2001-0x-Ti-S: 2.54 cm (1.0 inches) diameter, 9.91 cm (3.9 inches) length
Note: The length of the water level logger cable (CABLE-DR-xxx) can vary -0% to +3% +10 cm (3.9 inches) from the length ordered. The logger adds 38.8 cm (15.3) inches to the length of the cable ordered.
|
||||
Weight |
Top end (MX2001-TOP): Approximately 136 g (4.78 oz) in air
Stainless sensor end (MX2001-0x-SS-S): Approximately 141.4 g (4.99 oz) in air; approximately 93.4 g (3.3 oz) in fresh water Titanium sensor end (MX2001-0x-Ti-S): Approximately 80 g (2.83 oz) in air; approximately 37 g (1.3 oz) in fresh water |
||||
Wetted Materials |
Top end (MX2001-TOP): PVC housing, Polycarbonate end cap, Viton O-ring
Stainless sensor end (MX2001-0x-SS-S): Stainless steel housing, Viton and Buna-N O-rings, ceramic sensor in stainless steel end cap Titanium sensor end (MX2001-0x-Ti-S): Acetal housing, Viton and Buna-N O-rings, ceramic sensor in Titanium end cap Cable (CABLE-DR-XXX): Polycarbonate strain relief (top end connection), PVC end cap (sensor connection), nylon collar nuts, Viton O-rings, polyurethane jacket |
||||
Environmental Rating |
Top end: NEMA 6, IP67; Intermittent 100% RH
Sensor end: IP68 |
Reviews
There are no reviews yet.